Columnas Inogen One G5
  • Columnas Inogen One G5
  • Columnas Inogen One G5

Inogen One G5/Rove 6 Columns

Ref: RP-502
€145.00
Tax included
Quantity

The replacement of columns is easy and simple and can be done by the patient himself.

The steps for column replacement are as follows:

  1. Turn off the Inogen One G5/Rove 6 concentrator.
  2. Remove the battery from the concentrator.
  3. Position the concentrator so that you can see the column assembly.
  4. Unlock the column assembly by squeezing the unlock spring. Then pull the handle and remove the columns from the device. The columns are removed both at the same time in one structure.
  5. Then insert the new columns as far as they will go. The safety spring should spring back if the columns are properly engaged.
  6. Connect the Inogen One G5 to the mains, but KEEP THE DEVICE OFF.
  7. Press and hold the plus and minus buttons on the device for 5 seconds until "column replaced" appears on the display.
  8. Press the alert button and the display will show "column successfully replaced". Use the Inogen One G5 normally.
INOGEN
RP-502