Replacement Inogen One G4 concentrator columns.

Inogen One G4 concentrator columns should be replaced when maintenance is required and should not be taken as a standard practice.

The steps to replace the columns are as follows:

  1. Turn off the Inogen One® G4 concentrator by pressing the on/off button on the device.
  2. If necessary, remove the Inogen One® G4 concentrator from the carrying bag.
  3. Remove the battery from the Inogen One® G4 concentrator.
  4. Turn the Inogen One® G4 concentrator on its side so that you can see the underside. You can see the metal column assembly on the side of the device.
  5. Pull the lock button out.
  6. Holding the button open, pull the column handle and pull the column out of the device.
  7. Completely remove the columns from the Inogen One® G4. Both are removed in one piece.
  8. Remove the guards from the new column. Make sure that there is no dust or debris where they were located.
  9. Insert the column into the Inogen One® G4 concentrator. The edges of the columns should not be exposed; insert the column into the Inogen One G4 as soon as the protectors are removed.
  10. Push the column into the device so that it is fully seated in the Inogen One® G4 concentrator. The spring-loaded latch button should fully return to the closed position.
  11. Connect the AC power supply cord to the Inogen One® G4 and plug the other end into an electrical outlet. Do not turn on the Inogen One® G4 concentrator.
  12. Press and hold the plus and minus buttons for 5 seconds. The display will show the message "reset sieves". Release the button when this message has already been displayed on the screen.
  13. Press the alert button once ; the message "sieves reset successful" will appear on the display.
  14. Press the power button to turn the Inogen One® G4 on and use it as normal.